ST2515 Operation Manual
11
0 ADJ: Execute short correction. In this process, please ensure the test terminals have been
correctly shorted. Only by doing this, unexpected measurement errors could be avoided.
OVC (ON/OFF): ON is to use offset voltage compensation for measurement results. OFF is
not to.
Correction: Auto and Manual
Save data: Touch the screen or press the corresponding button to save the current
measurement results to a U disk. If no external U disk is found, this function will be inactive.
Once a U disk is found and this function is selected, the corresponding soft key will be
displayed as STOP SAVE and you can press this soft key again to terminate this operation.
NOTE:
SPEED: In SLOW2, SLOW1 or MED mode, the measurement result is a 6-digit number while FAST
is a 5-digit one.
SAVE: Once START SAVE is selected, you must press STOP SAVE to terminate and close the file
properly or the saved data will be lost.
SHORT: When this key is used to take short correction, the test fixture should be correctly
connected. If not, the deducted data will be wrong, and further it will cause result errors. The
correct connection is shown as the following figure.