ST2515 Operation Manual
17
3.4.4
TOOLS
CLEAR: Clear all parameters obtained from last statistic.
TRIGGER: Trigger the instrument to execute a measurement.
SAVE DATA: Touch this key to save the current measurement result (not all statistic results) to
an external U disk. If no U disk is found, this function will be disabled. Touch this key again to
stop saving.
3.4.5
FILE
1)
File Manage: Touch this key to enter into the file manage page.
2)
Screen copy: Touch this key to save the current screen page to an external U disk. If no U
disk is found, this function will be disabled.
3.5
<MEAS SETUP>
Touch the main menu. And then touch the soft key MEAS SETUP to enter into the MEAS SETUP
page shown as figure 3-5.
Figure 3-5 Measurement Setup
On this page, the following parameter can set:
FUNC R-T: Touch this key to select the required parameter.
RANGE AUTO: Touch this key, the following soft keys will be displayed: AUTO (automatically
set the resistance range), HOLD (lock the resistance range), ↑ (increase the range), ↓
(decrease the range). Touch the corresponding soft key to set the range mode.
TRIG INT: Touch this key, the following soft keys will be displayed in the soft key zone: INT
(auto measurement mode), MAN (manual measurement mode), EXT (external trigger mode),