17
After the test is finished, the value can be recovered by pressing the
key.
3.10. Built-in printer
In order to enable the printing function press key. Measured values will
be printed each 15 seconds, and the Dielectric Absorption Index and
Polarization Index will be printed after 1 minute and 10 minutes
respectively. Printing may be started or stopped at any time during the
test. However, it is convenient to turn the printer on before starting the
test in order to print it complete, including the heading.
3.11. Battery status check
During the measurement, it is possible to check the battery status. The
key must be pressed. If the battery charge is enough, it will read “BAT
OK”. If the charge is not enough, the message will be “LOW BAT”, and
the battery should be recharged. The analogue bargraph will give an
approximate idea of the remaining charge percentage (at least 20% is
required for a normal operation).
When the battery charge reaches the normal operation minimum value,
the message LO BAT appears automatically in the area where the test
tension value is indicates, alternating it every 1 second.