12
It has a real time clock with date, hours and minutes indication, to make
identification of tests recorded in paper or in memory easier.
Test number
Tests are automatically numbered by the equipment to make their
identification easier. The test number is printed at the beginning of each
test and it is stored in memory.
Model and serial number
At the beginning of each test, the equipment model, as well its serial
number, are registered, making it possible to relate the obtained results
with their respective equipment Calibration Certificate.
3. Tests definition
The
MD-5060x
is an extremely versatile equipment, which allows to carry
out several kinds of insulation tests automatically, recording all the results
in its internal memory or printing them. It is necessary to define the tests
properly, selecting its parameters before starting the test:
Test voltage
Equipment operation mode
Minimum resistance limit, for the “Pass-fail test”
These three parameters shall be defined before starting the test. The test
voltage is the only parameter that can be modified during the test. The
keys makes it possible to enter the different equipment
programming modes. The
(decrease)
and
(increase) keys select the
value. The functions of the
keys are mutually exclusive. Selecting
one of these keys implies releasing any other selected key.