21
Application Note 32 explains the usage of Guard terminal for
minimizing the parasite resistance effect, whose influence one
intends to minimize.
4. Turn the equipment by using the On/Off switch.
5. The following message will appear on the display:
6. The
PRESS START
message will appear immediately.
7. In order to make repetitive tests performance easier, when the
equipment is turned on, the test voltage will be the same voltage
selected for the last performed test, meanwhile the equipment will be
in the 500 V steps voltage test programming mode ( key led
flashing). To modify this voltage, use the and keys or the fast
selection
keys. For a voltage test fine adjustment (in 100 V
steps) press the key. In this condition, the and keys make it
possible to decrease or increase the 100 V steps test voltage to
select 500, 600, 700, ... 5000 Vdc.
8. To leave the voltage test programming mode just press the activated
key again in order to deactivate it (the associated led will stop
flashing). Now, the voltage selected value will be recorded in the
equipment RAM memory for future tests.