9. Press key. The
high voltage led
turns on immediately,
indicating that the equipment internal generator is applying
voltage to the element that is being tested. If, during the
test, it is necessary to change the test voltage, item 7
sequence shall be repeated.
10. For a few seconds, the self-scale smart system will search for the
most convenient range for the value being measured. Meanwhile, the
display will show the message:
11. Then, the display will show the test number, the selected voltage
value and it will start the elapsed time count. If the measured value is
within the device scope, the test number indication will provide place
for the resistance value indication and its corresponding unit, and it
will start the analogue bargraph indication.
12. If, for example, the measured value is 601 M
Ω with a selected
voltage of 1000 V the display will initially remain on during a few
seconds, as it is shown below, informing that test 12 was started.
13. After a few seconds, in the same example, the display will indicate
that the measured resistance value is 601 M
Ω, as it is shown: