5-24-3. Y SYNC Cancel Adjustment
Equipment
:
Oscilloscope
Preparation :
• TEST 2 button MSU operation panel
→
ON (Lights)
Test Point
:
TP1 (GND:E1)/MD-103 (C-3)
Adjustment Point :
1
RV2 (SYNC CANCEL)/
MD-103 (A-3)
Adjustment Procedures:
Adjust
1
RV2 so that the H SYNC portion is flat.
5-24-4. Y/SKIN DC Balance Adjustment
Equipment
:
Oscilloscope (LIMITER
→
OFF)
Test Point
:
TP2 (GND:E1)/MD-103 (C-2)
Trigger
:
SYNC OUTPUT connector/
CCU rear panel
Adjustment Points :
1
RV3 (Y DC BAL)/MD-103 (B-3)
1
RV4 (SKIN DC BAL)/MD-103 (C-3)
Adjustment Procedure :
Adjust
1
RV3 and
1
RV4 alternately until the V SYNC
portions are flat.
V rate
MD-103 BOARD (A SIDE)
MD
PROMPT
GENLOCK
A
1
2
3
4
B
C
D
E
F
G
RV2
RV3
TP2
RV4
TP1
E1
H rate
Flat
Resetting after Adjustment:
• TEST 2 button MSU operation panel
→
OFF (Goes out)
5-24. Video Modulation/Demodulation Adjustment
5-24-3. Y SYNC Cancel Adjustment
5-24-4. Y/SKIN DC Balance Adjustment