Fraxel
®
1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Laser System
Part Number 10-05170 Rev. O
Page 80 of 92
9.0 Fraxel Laser Systems Specifications
9.1 Treatment
Beam
Invisible Laser Radiation
Invisible Laser Radiation
Product
Fraxel 1550
Fraxel 1927 and Fraxel
DUAL 1550/1927
Equipment
classification
IEC60601-1
Class I
Class I
CDRH Laser
Classification
Class 4
Class 4
IEC 60825
Classification
Class 4
Class 4
Wavelength
1550 min:1520
max:1560
1550 min:1520
max:1560
1927 min:1905
max:1945
Maximum Power
30 W
30 W / 12 W
Maximum Pulse
Energy
70 mJ
70 mJ / 20 mJ
Maximum Pulsewidth
10 ms
10 ms
Pulse Repetition Rate
0 – 3kHz
0 – 3kHz
9.2 Console
Dimensions / in
19.25w x 17.75h x 17.5d
(/cm)
48.9w x 45.1h x 44.45d
Weight
~ 55 lbs
Warning:
SEVERE DAMAGE TO THE UNIT MAY RESULT FROM IMMEDIATE POWER UP
FOLLOWING MOVEMENT OF THE UNIT FROM A COLD ENVIRONMENT INTO A WARM
ENVIRONMENT. THIS CONDITION WILL OCCUR IF FOR EXAMPLE THE UNIT IS
MOVED FROM LOCATION TO LOCATION ON A DAILY BASIS AND IS STORED IN A
SHIPPING CONTAINER, AUTOMOBILE, OR TRUCK OVERNIGHT.
THE UNIT SHOULD BE ALLOWED TO EQUILIBRATE WITH ITS SURROUNDINGS FOR A
MINIMUM OF 90 MINUTES FOLLOWING A CHANGE OF ENVIRONMENT. F AILURE TO
OBSERVE THESE PRECAUTIONS WILL RESULT IN THE WARRANTY OF THE UNIT
BEING VOIDED.