Fraxel
®
1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Laser System
Part Number 10-05170 Rev. O
Page 18 of 92
2.3 System
State
Descriptions
This section provides a general overview of the Fraxel Laser Systems operation.
These laser systems have specific operating requirements. It is the responsibility of
the physician to fulfill these requirements. Failure to do so can result in incorrect
and/or intermittent operation or damage to the laser.
While powering up, the system will briefly display an introductory splash screen.
2.3.1
System Test Mode
Upon completion of the system power-up, the system will enter the System Test mode.
Please follow the instructions as they appear on the touch screen display. The initial system
test screen will direct the placement of the handpiece into the calibration port.
In order to perform system testing, remove the tip from the handpiece. Place the tip in a
clean, dust-free container while the testing is in progress. Then, proceed by placing the
handpiece in the calibration port located on the top right hand side of the laser console. The
touch screen will indicate when testing has initiated.