Warning:
Do not allow the chip inside the tip to come into contact with any liquid such as the
Cidex
®
OPA solution, or water.
Do not allow the metal part of the handpiece to come into prolonged contact with any
liquid such as the Cidex
®
OPA solution.
Pre-cleaning Procedure
After tip use, remove and discard the tip rollers. Wipe off the outside of the tip with 70%
isopropyl alcohol (IPA) wipes until there is no visual sign of soil. Tips can then be placed, but
not submerged, in the Cidex solution (per the Cidex OPA solution instructions).
DO NOT
submerge the tip completely.
DO NOT
allow the cleaning solution to come in contact with
the chip inside the tip.
If desired, pre-cleaning can be done with a solution of an enzymatic detergent, e.g. Enzol
®
enzymatic detergent, (per the manufacturer’s recommendations). Pre-clean devices by
immersing the soiled portion of the tips in the prepared detergent.
During cleaning or disinfection,
DO NOT
allow any liquid to come into contact with the chip
inside the tip.
Rinse the part of the tip exposed to either IPA or enzymatic detergent with water for a
minimum of 30 seconds, being careful not to expose the chip to any liquid.
High Level Disinfection Procedure
Pour the Cidex
®
OPA solution into a container and immerse the base of the tip (where the
white rollers attach). Ensure that the Cidex
®
OPA solution completely covers the base of the
tip. Allow the tip to soak in the disinfectant for 12 minutes (or per Cidex OPA instructions for
use).
Following disinfection with Cidex
®
OPA solution, rinse the tip with water and then dry the tip
thoroughly and attach new unused rollers before use.
The rollers used in the Fraxel Cool Roller Tip, are for single patient use only. Do not
reuse. Multiple patient use may create the potential risk of cross contamination of
biological agents from one patient to another.
More details on the Enzol
®
and Cidex
®
OPA products can be found at:
http://www.aspjj.com/us/products/enzol/features-and-benefits
and
http://www.aspjj.com/us/products/cidex-opa/features-and-benefits