Fraxel
®
1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Laser System
Part Number 10-05170 Rev. O
Page 34 of 92
2.4.3
Measurement Process
The Operator must establish contact between the handpiece and the surface of the
treatment region. Vertical and horizontal measurements must be completed as directed by
the measurement touch screen display.
After completion of the measurements for a particular treatment area, the Operator may
select an additional treatment region for measurement. In order to do so, the Operator must
select “Change Treatment Region” and must select a new treatment location. After
completion of vertical and horizontal measurements for the selected region(s), the Operator
must press “Done”. Upon completion of the requested measurements, the system will
provide the value for surface area (cm
2
).