9.3 Laser Safety Eyewear
Refer to “Optical Safety” – Section 3.0 of this document
9.4 Cart
Dimensions / in
20w x 30h x 22d
(/cm)
50.8w x 76.2h x 55.9d
Weight <100
lbs
9.5 Operating
Environment
Maximum Recommended Ambient Air
Temperature for Treatment
35°C / 95°F
Operating Temperature Range
15 to 40°C, 59 to 104°F
Humidity
30 – 75% Non-condensing
Altitude
< 10000 feet / 3075 meters
9.6 Shipping and Storage (Non-Operational)
Temperature Range
-15 to 45°C, 5 to 113°F
Humidity
30 – 75% Non-condensing
Altitude
< 45000 feet / 13846 meters
9.7 Compatible
Delivery
Devices
Only use original Fraxel 1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Disposable Tips.
Contact your local authorized distributor of Solta Medical lasers for more information.
9.8 Electromagnetic compatibility and immunity
This guidance and manufacturer's declaration information pertains to the following Solta
Medical products: Fraxel DUAL 1550/1927 Laser System
Special precautions concerning electromagnetic compatibility (EMC) must be taken for all
medical electrical equipment.
All medical electrical equipment must be installed and put into service in accordance with
the EMC information provided in this document.
Portable and mobile RF communications equipment can affect the behavior of medical
electrical equipment.
The Fraxel DUAL 1550/1927 Laser System comply with all applicable and required
standards for electromagnetic interference:
It does not normally affect nearby equipment and devices.