11.4 Environmental
The temperature in the laser room should ideally remain between 65º to 80º F, (18 – 27°C)
both for patient comfort and for optimum operation of the laser. Patients may feel more
discomfort if the temperature of the room is excessively high. A maximum ambient
operating temperature of 104°F / 40ºC is allowed. The cooling system for the laser is forced
air, and no other external utilities are required.
The Fraxel 1550 and Fraxel DUAL 1550/1927 Laser Systems will emit a maximum of 600W
/ 2050BTUh into the environment around the console.
11.5 Electrical
The Fraxel 1550 and Fraxel DUAL 1550/1927 Laser Systems contain a universal, medical
grade, power-factor corrected power supply. The line voltage and power switch amperage
requirements are given in the specifications section of the manual. Solta recommends a
dedicated circuit for the laser system. Place the power cord where it cannot become a
hazard, for example by being tripped over or having sharp objects dropped on it.
The worst-case current draw is anticipated to be:
100-240V
50/60Hz 5
2.5A
Warning:
The operator is responsible for ensuring that the proper electrical specifications are met at
the installation site.
Warning:
Laser equipment not in use should be protected against unauthorized use by removing the
key from the key switch. Keep the key in a designated place accessible only to designated
and trained personnel. Moving the Laser
11.5.1
Shut down the laser and then remove the power cord from the
wall and from the power switch housing in the laser.
11.5.2
Remove the footswitch cable plug from the rear of the system
and carefully wrap the footswitch cable around the footswitch
housing. Store the footswitch in the basket in the rear of the
cart. If the external interlock is connected, remove this plug.
11.5.3
Lock the handpiece umbilical management arm prior to
movement. Unlock the cart wheels.