Warning:
All personnel within the Nominal Ocular Hazard Distance (NOHD) must be wearing
approved eye protection at this point. Failure to wear appropriate eyewear increases the
risk of serious eye injury.
Warning:
Laser equipment not in use should be protected against unauthorized operation by removing
the key from the key switch. Keep the key in a designated place accessible only to
authorized and trained personnel to avoid unauthorized use by untrained personnel and the
possibility of injury or fire.
Warning:
Do not attempt to coil the umbilical beyond the resistance point of the armoring. Do not use
if the internal cables and wires are visible through holes or gaps in the umbilical armoring as
this may cause inadvertent exposure of laser energy to persons in the room.
Warning:
If it is necessary to manually lift the Laser System console and cart, or to separate the
console from the cart, two people must perform the lifting operation. Failure to follow this
directive may result in serious injury or damage to the unit.
Warning:
SEVERE DAMAGE TO THE UNIT WILL RESULT FROM IMMEDIATE POWER UP
FOLLOWING MOVEMENT OF THE UNIT FROM A COLD ENVIRONMENT INTO A WARM
ENVIRONMENT. THIS CONDITION WILL OCCUR IF FOR EXAMPLE THE UNIT IS
MOVED FROM LOCATION TO LOCATION ON A DAILY BASIS AND IS STORED IN A
SHIPPING CONTAINER, AUTOMOBILE, OR TRUCK OVERNIGHT.
THE UNIT MUST BE ALLOWED TO EQUILIBRATE WITH ITS SURROUNDINGS FOR A
MINIMUM OF 90 MINUTES FOLLOWING A CHANGE OF ENVIRONMENT. FAILURE TO
OBSERVE THESE PRECAUTIONS WILL RESULT IN THE WARRANTY OF THE UNIT
BEING VOIDED.
NOTE:
NEVER BLOCK THE COOLING SYSTEM VENTS. ENSURE ADEQUATE SPACE
AROUND THE UNIT AT ALL TIMES TO FACILITATE SYSTEM COOLING.
CAUTION: USE OF CONTROLS, ADJUSTMENTS, OR PROCEDURES OTHER THAN
THOSE SPECIFIED MAY RESULT IN HAZARDOUS ELECTROMAGNETIC RADIATION
EXPOSURE WHICH MAY RESULT IN INJURY, OR DAMAGE TO OTHER EQUIPMENT.