Fraxel
®
1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Laser System
Part Number 10-05170 Rev. O
Page 9 of 92
1.2 Device
Overview
The Fraxel Laser Systems are designed for cutaneous laser treatment. These devices are
part of the family of Solta Medical products utilizing the Fractional Photothermolysis
principle.
The Fraxel 1550 consists of a single infra-red fiber laser with wavelength of 1550 nm.
The Fraxel 1927 consists of a single infra-red Thulium fiber laser with wavelength of 1927
nm.
The Fraxel DUAL 1550/1927 Laser System has two infra-red fiber lasers with wavelengths
of 1550 nm and 1927 nm. The two wavelength option allows for targeted treatment of
different tissue depths. The 1550 nm laser coagulates the epidermis and dermis with up to
1.6 mm depth of penetration. The 1927 nm laser coagulates epidermis and superficial
dermis with up to 0.3 mm depth of penetration. The laser sources are controlled by an
embedded processor. The solid state design of the Fraxel 1550, Fraxel 1927 and Fraxel
DUAL 1550/1927 Laser Systems typically result in limited maintenance and minimal utilities
requirements. The output of the fiber laser is focused into the skin using lenses contained
within the handpiece. Computer controlled motors within the handpiece and the software
control architecture direct the laser system to deliver focused spots in an evenly spaced
treatment pattern. These advanced features allow for precisely controlled delivery of the
fractional laser treatment.
Interchangeable tips allow for treatment of both small and large areas. The small tip design
is intended to allow for treatment of smaller areas (not available with all systems).
This manual is intended to help you get the most from your system as you make the Fraxel
1550, Fraxel 1927 and Fraxel DUAL 1550/1927 Laser Systems an integral part of your
practice. It is our intention that the information provided assists the operator in successful
use of this product. This manual is not intended to substitute for training by authorized Solta
Medical personnel prior to treatment of patients. If you have additional questions, please
contact your local Solta Customer Service representative directly.
Warning:
The Fraxel Laser Systems generate an intense beam of laser radiation which may cause
injury if not used properly. It is essential that this manual be read thoroughly and
understood in its entirety before operation.
1.3 Introduction to the System
The Fraxel 1550 and Fraxel 1927 systems both have a single fiber laser source, and the
Fraxel DUAL 1550/1927 Laser System has two fiber laser sources. The fiber laser controls
are contained in a single console. The console is electrically connected to the facility power
source. Laser energy produced by the unit is delivered to the tissue through the handpiece.
Contact with the tissue is maintained by removable disposable contact tips which attach to
the handpiece.
The features of the Fraxel Laser Systems are shown below: