40
•
Low Priority Alarm:
This alarm
does not stop the pump (delivery
continues). The alarm is accompanied
by a
blue
screen and persists for
5 seconds (or more with some alarms)
unless it is acknowledged by a key
press or until the condition that
triggered the alarm goes away.
•
Informational priority message:
These messages do not stop the
pump. They appear in the status bar for 5 seconds, are almost
always silent, and require no acknowledgement. Examples include
“Cassette Locked” and “Cassette Unlocked”.