20
Removing a Cassette
The medication reservoir should be changed before new infusions,
and when you receive reservoir volume empty or low alarms.
Note:
You should always have the supplies necessary to change
your CADD
™
medication cassette reservoir or CADD
®
administration set, as instructed by your health care provider.
WARNING: Always clamp the tubing before removing the
cassette from the pump. Removing the cassette without
closing the clamp could potentially cause unregulated gravity
infusion, which could result in serious injury or death.
To remove a cassette:
1. Make sure the pump is stopped before removing the cassette.
For instructions, see “Stopping the Pump During an
2. Close the tubing clamp.
3. If the cassette is locked, insert the pump key
into the cassette/keypad lock and turn it
counter-clockwise to the unlocked position.
4. Push down on the
cassette latch until the
cassette detaches.
5. Remove and discard the CADD
™
medication cassette
reservoir or CADD
®
administration set as instructed by your
health care provider.
Summary of Contents for cadd-solis VIP 2120