Cautions
•
Do not operate the pump at temperatures below 2°C (36°F) or
above 40°C (104°F) to avoid damaging the electronic circuitry.
•
Do not store the pump at temperatures below –20°C (–4°F) or
above 60°C (140°F) to avoid damaging the electronic circuitry.
Do not store the pump with a CADD
™
medication cassette
reservoir or CADD
®
administration set attached.
•
Do not expose the pump to humidity levels below 20% or above
90% relative humidity to avoid damaging the electronic circuitry.
•
Inspect the AA batteries for damage or wear to the metal or
plastic insulation prior to use, or after the pump has been
dropped or hit. Replace the batteries if any damage is noted.
•
Do not store the pump for prolonged periods with the batteries
installed. Battery leakage could damage the pump.
•
If the power up results in an error message indicating that the
protocol library was lost, do not proceed with using the pump.
Contact your health care provider.
•
If delivery of an infusion is affected by a time or date change, an
alarm message appears and must be confirmed.
•
Do not immerse the pump in cleaning fluid or water. Do not
allow solution to soak into the pump, accumulate on the keypad,
or enter the battery compartment, USB port, remote dose cord
jack, or power jack areas. Moisture buildup inside the pump may
damage the pump.
•
Do not clean the pump with acetone, other plastic solvents, or
abrasive cleaners, as damage to the pump may occur.
•
The pump should
not
be directly irradiated by therapeutic levels
of ionizing radiation due to the risk of permanent damage to the
electronic circuitry. The best procedure to follow is to remove the
pump during therapeutic radiation sessions or diagnostic levels
of radiographic and fluoroscopic radiation. If the pump must
remain in the vicinity during a diagnostic or therapy session, it