30
If your health care provider instructed you to change your IV bag
without changing the CADD® administration set, follow the steps
below.
To reset the reservoir when changing an IV bag only:
1. Stop the pump.
2. Remove the used IV bag from the administration set and
attach the new IV bag to the administration set, as instructed
by your health care provider.
3. In the
Tasks
menu, press
U
or
D
to highlight
Reset Reservoir Volume,
and press
S
.
4. The pump displays a question asking
if you wish to reset the reservoir
volume to the default amount for your
therapy (in this example, 265 mL).
Select
Yes
to reset the volume.
5. If you selected to reset the reservoir
volume, the pump shows a
Saving…
screen.
6. Press the stop/start
9
key to start
the pump.