type of drug being administered, could result in death or serious
injury.
•
If the pump is dropped or hit, the battery door may become
broken or damaged. Do not use the pump if the battery door
is damaged because the batteries will not be properly secured.
This may result in loss of power and nondelivery of drug, and
depending on the type of drug being administered, could result
in death or serious injury.
To install batteries or the rechargeable battery pack:
1. Make sure the pump is stopped and powered off.
2. Using your fingers or a coin, turn the knob
on the battery door counterclockwise to
open the battery door.
3. Hold the pump at an angle to
remove the old batteries and insert
4 new AA batteries. Match the
+
and
–
markings on the batteries
with the markings on the pump.
OR:
If using a rechargeable battery
pack, insert it as shown.
4. Close the battery door and turn the knob on
the battery door clockwise to lock.
Note:
•
If the batteries are installed backwards, the
pump will not power on. If the pump does
not power on, check the batteries, making
sure to match the
+
and
–
markings on the
batteries to the pump.
•
Battery life may vary depending on the amount and rate of
medication delivered, battery age, temperature, and active display
time and backlight intensity (increasing backlight intensity
shortens battery life).
•
Battery power is quickly depleted at temperatures below 10°C (50°F).