HRX-OM-M090
Chapter 5 Display and setting of various functions
HRS Series
5.19
Communication function
5-65
42.
Select contact input signal 2 delay timer of reading from the table below with [
▲
] key or
[
▼
] key, and confirm by pressing “SEL”.
Table 5.19-22
List of set value
Set value
Explanation
Initial value
(Default setting)
Setting and checking are not available unless
contact input singal 2 is external switch signal
input (N.O. type or N.C. type).
to
Setting of contact input signal 2 delay timer of
reading.
Set range is 0 to 300 sec.
Contact input signal 2 OFF detection timer
Setting
and checking
43.
Press the [SEL] key once.
The set screen of contact input signal 2 OFF detection timer is displayed on the digital
display.
PV
SV
44.
Select contact input signal 2 OFF detection timer from the table below with [
▲
] key or
[
▼
] key, and confirm by pressing “SEL”.
Table 5.19-23
List of set value
Set value
Explanation
Initial value
(Default setting)
Setting and checking are not available unless
contact input singal 2 is external switch signal
input (N.O. type or N.C. type).
to
Setting of contact input signal 2 OFF
detection timer
Set range is 0 to 10sec.
Contact output signal 1 function
Setting
and checking
45.
Press the [SEL] key once.
The set screen of contact output signal 1 function
is displayed on the digital display.
PV
SV