-62-
No.PF
※※
-OMM0005-N
4. Input of security code (3 digit setting)
The hundreds digit starts flashing.
Press the
or
button to change the set value.
Press the
button to make the next value to the right flash.
(If the
button is pressed on the far right digit, the hundreds digit
will flash.)
After the setting is complete, press the
button continuously for
1 second or longer.
(If an operation is not performed for 30 seconds during input or
change of the security code, it will return to measurement mode
with [LoC] status.)
If the password is wrong, [FAL] will be displayed on the sub
screen. In this case, input the password again.
If the wrong security code is entered 3 times, it will return to
measurement mode.
[UnLoC] is displayed on the sub screen.
Press the
button to complete the unlocking operation.
●
Change of security code
In the default setting, security code entry is set to [000], but this can be changed to any number.
<Operation>
1. After the key lock setting is complete, perform all four steps in the key unlocking procedure.
(Refer to Key-lock function on page
.)
2. After the security code is entered and the indication changes to [UnLoC] on the sub screen, press and
hold the
and buttons simultaneously for 5 seconds or longer.
[000] is displayed and the new security code should be input.
For how to input the security code, refer to "4" Input of security code on page
Press the
button for 1 second or longer.
The new security code is displayed on the sub screen.
Press the
button for 1 second or longer.
The change of security code is complete.
The status remains [UnLoC] after the change is completed, so perform the key lock setting again to change
it to [LoC].
Summary of Contents for PF3W7 Series
Page 12: ...11 No PF OMM0005 N Model Indication and How to Order Metal piping...
Page 13: ...12 No PF OMM0005 N Polyvinyl chloride piping...
Page 45: ...44 No PF OMM0005 N List of output modes for fluid temperature...
Page 48: ...47 No PF OMM0005 N Example of display of sub screen...
Page 49: ...48 No PF OMM0005 N Example of display of sub screen continued...
Page 78: ...77 No PF OMM0005 N PF3W711 Vinyl chloride piping PF3W721 Vinyl chloride piping...