![SMAR HI302 User Manual Download Page 91](http://html1.mh-extra.com/html/smar/hi302/hi302_user-manual_1294915091.webp)
Appendix C
C.13
HVT ALLOCATION MAP FOR SMAR TT301 V2.04
Index
Parameter Name
HART Variable Name
HART Command
Description
16.9
FLOAT_ARRAY_1[9]
PROPORTIONAL_FACTOR(KP)
142,143
w/r
Proportional Factor Value
16.10
FLOAT_ARRAY_1[10]
INTEGRAL_TIME(TR)
142,144
w/r
Integral Time Value
16.11
FLOAT_ARRAY_1[11]
DERIVATIVE_TIME(TD)
142,145
w/r
Derivative Time Value
16.12
FLOAT_ARRAY_1[12]
NON_LINEAR_FATOR(KNL)
142
r
Non Linear Factor Value (always 0)
16.13
FLOAT_ARRAY_1[13]
DERIVATIVE_FACTOR(DG)
142
r
Derivative Factor Value (always 10)
16.14
FLOAT_ARRAY_1[14]
MANIPULATED_VARIABLE_HIGH_LIMIT
148,149
w/r
MV maximum limit
16.15
FLOAT_ARRAY_1[15]
MANIPULATED_VARIABLE_LOW_LIMIT
148,149
w/r
MV minimum limit
16.16
FLOAT_ARRAY_1[16]
MANIPULATED_VARIABLE_MAXIMUM
148,149
w/r
MV maximum rate of changing
16.17
FLOAT_ARRAY_1[17]
POWER_ON_SETPOINT
150,151
w/r
Power_On SP
16.18
FLOAT_ARRAY_1[18]
POWER_ON_MANUAL_OUTPUT
150,151
w/r
Fail safe value
16.19
FLOAT_ARRAY_1[19]
SETPOINT_TIME
152,153,184
w/r
SP time
17.2
FLOAT_ARRAY_2[2]
UPPER_RANGE_VALUE
158
w
Save Upper Range Value to EEPROM
17.3
FLOAT_ARRAY_2[3]
LOWER_RANGE_VALUE
158
w
Save Lower Range Value to EEPROM
17.4
FLOAT_ARRAY_2[4]
ALARME_1_VALUE
159,160
w/r
Alarm 1 value limit
17.5
FLOAT_ARRAY_2[5]
ALARME_2_VALUE
159,160
w/r
Alarm 2 value limit
17.6
FLOAT_ARRAY_2[6]
SPECIAL_SENSOR_UPPER_RANGE_LI
MIT
177,178
w/r
Special sensor upper range limit
17.7
FLOAT_ARRAY_2[7]
SPECIAL_SENSOR_LOWER_RANGE_LI
MIT
177,178
w/r
Special sensor lower range limit
17.8
FLOAT_ARRAY_2[8]
SPECIAL_SENSOR_MINIMUM_SPAN
177,178
w/r
Special sensor minimum span limit
17.9
FLOAT_ARRAY_2[9]
SPECIAL_SENSOR_HIGH_LIMIT
177,178
w/r
Special sensor high limit
17.10
FLOAT_ARRAY_2[10]
SPECIAL_SENSOR_LOW_LIMIT
177,178
w/r
Special sensor low limit
17.11
FLOAT_ARRAY_2[11]
SPECIAL_SENSOR_OHMS_MV_UPPER_
LIMIT
179,180
w/r
Special sensor mv or ohm upper limit
17.12
FLOAT_ARRAY_2[12]
SPECIAL_SENSOR_OHMS_MV_LOWER
_LIMIT
179,180
w/r
Special sensor mv or ohm lower limit
17.13
FLOAT_ARRAY_2[13]
OUTPUT_VARIABLE
184
r
Output Variable (OUT) Value
17.14
FLOAT_ARRAY_2[14]
MEAS_PV_CURR_LEVEL_GAIN
46
w
Set value of the Trim Primary Variable Current Dac Gain
17.15
FLOAT_ARRAY_2[15]
CAL_POINT_ZERO
134
w
Zero user calibration point
17.16
FLOAT_ARRAY_2[16]
SP_CURVE_X1
156,157
w/r
Coordinate X of set point curve index 1
17.17
FLOAT_ARRAY_2[17]
SP_CURVE_X2
156,157
w/r
Coordinate X of set point curve index 2
17.18
FLOAT_ARRAY_2[18]
SP_CURVE_X3
156,157
w/r
Coordinate X of set point curve index 3
17.19
FLOAT_ARRAY_2[19]
SP_CURVE_X4
156,157
w/r
Coordinate X of set point curve index 4
17.20
FLOAT_ARRAY_2[20]
SP_CURVE_X5
156,157
w/r
Coordinate X of set point curve index 5
18.1
FLOAT_ARRAY_3[1]
SP_CURVE_X6
156,157
w/r
Coordinate X of set point curve index 6
Summary of Contents for HI302
Page 1: ...AGO 02 FOUNDATION H I 3 2 F M E...
Page 4: ...HI302 User s Manual IV...
Page 10: ...General Vision 1 4...
Page 20: ...Installation 2 10...
Page 46: ...Operation 5 12...
Page 57: ...HI302 User s Manual 7 5...
Page 59: ...HI302 User s Manual 7 7...
Page 60: ...An Example of HI302 Usage 7 8...
Page 64: ...Troubleshooting 8 4...