Appendix A
A.1
HCFG - HART Configuration Transducer Block
Index
Parameter
Data Type
Valid Range / Options
Default Value
Units Store / Mode
Description
0
BLOCK_STRUCTURE
DS-64
NA
S
1
ST_REV
Unsigned16
0
None
S / RO
2
TAG_DESC
OctString(32)
Spaces
NA
S
3
STRATEGY
Unsigned16
0
None
S
4
ALERT_KEY
Unsigned8
1 to 255
0
None
S
5
MODE_BLK
DS-69
O/S
NA
S
See Mode Parameter
6
BLK_ERR
Bitstring(2)
None
D / RO
7
FIRMWARE_VERSION
VisibleString(32)
0
NA
S / RO
This field is useful for information and diagnostic purposes.
8
COMM_BEHAVIOR
Boolean
TRUE:
Autonomous
FALSE:
Bypass
TRUE
NA
S
Autonomous means the normal behavior where the database is built
automatically. If set to Bypass the device depends on external applications
to send HART commands.
9
COMM_ENABLE
Boolean
TRUE:
Enabled
FALSE:
Disabled
FALSE
NA
S
FALSE disables HART Communication on all the channels. Set this
parameter to FALSE before make any configuration change and set it back
TRUE again in order to rebuild and validate the new configuration data
base.
10
CHANNEL_ACTIVE
Unsigned8[8]
0x00:
NO
0x01:
1
0x02:
2
0x0F:
15
0x00: NO
NA
D / RO
If the correspondent channel has at least one HART device configured,
the value is set to the number of devices on that channel. If
NO,
there are
no devices configured for the channel and there will no scan this channel
until there is at least one device.
11
SIMUL_COMM_ENABLE
Boolean
TRUE:
Enabled
FALSE:
Disabled
TRUE
NA
S
Set TRUE if the simultaneous communications over all the 8 channels is
allowed.
12
MASTER_TYPE
Unsigned8[8]
0x01
: Primary
0x00
: Secondary
0x01: Primary
None
S
HART Master Type.
13
RETRIES
Unsigned8[8]
3 to 10
3
None
S
Number of retries if slave does not respond before slave timeout or if any
error is received. HART Data Link Layer parameter.
14
MASTER_SYNCHRONIZED
Boolean[8]
TRUE:
Synchronized.
FALSE:
Not synchronized
FALSE
NA
D / RO
If FALSE, the Master Channel can't synchronize the state machine for
transmission.
15
CHANNEL_MODE
Unsigned8[8]
0x00:
Normal
0x01:
Burst Mode
0x00: Normal
None
D / RO
This parameter shows if any burst mode device was detected on the
respective channel.
16
MASTER_STATE
Unsigned8[8]
0x00:
Watching
0x01:
Enabled
0x02:
Using
0
None
D / RO
It's the Master State Machine behavior at each moment.
17
REQUEST_COUNTER
Unsigned32[8]
0
None
D / RO
Counts the number of requests made to all devices on that channel.
18
RETRIES_COUNTER
Unsigned32[8]
0
None
D / RO
Counts the number of retries to all devices on each channel.
19
INVALID_SOM
Unsigned32[8]
0
None
D / RO
Counts the number of invalid Start Of Messages captured in that channel.
20
INVALID_RX_FRAMES
Unsigned32[8]
0
None
D / RO
It totalizes the number of INVALID frames received by each channel,
whichever the error.
21
VALID_RX_FRAMES
Unsigned32[8]
0
None
D / RO
It totalizes the number of VALID frames received by each channel.
Summary of Contents for HI302
Page 1: ...AGO 02 FOUNDATION H I 3 2 F M E...
Page 4: ...HI302 User s Manual IV...
Page 10: ...General Vision 1 4...
Page 20: ...Installation 2 10...
Page 46: ...Operation 5 12...
Page 57: ...HI302 User s Manual 7 5...
Page 59: ...HI302 User s Manual 7 7...
Page 60: ...An Example of HI302 Usage 7 8...
Page 64: ...Troubleshooting 8 4...