HI302 – User’s Manual
C.12
HVT ALLOCATION MAP FOR SMAR TT301 V2.04
Index
Parameter Name
HART Variable Name
HART Command
Description
12.17
U8B_ARRAY_2[17]
ALARM_2_ACKNOWLEDGE
159,161
w/r
Acknowledge alarm 2
12.18
U8B_ARRAY_2[18]
FAIL_SAFE_MODE
162,163
w/r
Fail_safe mode (low or high)
12.19
U8B_ARRAY_2[19]
PV_DISPLAY
164,165
w/r
PV for Display indication
12.20
U8B_ARRAY_2[20]
SV_DISPLAY
164,165
w/r
SV for Display indication
13.1
U8B_ARRAY_3[1]
COUNT_CHANGE_ZERO_SPAN
166
r
Number of time the zero and span trim was done
13.2
U8B_ARRAY_3[2]
COUNT_CHANGE_FUNCTION
166
r
Number of time the function was changed
13.3
U8B_ARRAY_3[3]
COUNT_CHANGE_TRIM_4MA
166
r
Number of times the 4 mA trim was done
13.4
U8B_ARRAY_3[4]
COUNT_CHANGE_TRIM_20MA
166
r
Number of time the 20 mA trim was done
13.5
U8B_ARRAY_3[5]
COUNT_CHANGE_TRIM_USER
166
r
Number of time the user trim was done
13.6
U8B_ARRAY_3[6]
COUNT_CHANGE_BURNOUT
166
r
Number of times the Burnout action was changed
13.7
U8B_ARRAY_3[7]
COUNT_CHANGE_SENSOR
166
r
Number of time the sensor was changed
13.8
U8B_ARRAY_3[8]
COUNT_CHANGE_OPERATION_MODE
166
r
Number of time the mode was changed
13.9
U8B_ARRAY_3[9]
COUNT_CHANGE_LOCAL_ADJUST
166
r
Number of time the local adjust protection was changed
13.10
U8B_ARRAY_3[10]
COUNT_CHANGE_COMMUNICATION
166
r
Number of time the communication protection was changed
13.11
U8B_ARRAY_3[11]
COUNT_CHANGE_MULTIDROP
166
r
Number of time the multidrop address was changed
13.12
U8B_ARRAY_3[12]
COUNT_CHANGE_PASSWORD
166
r
Number of time the password was changed
13.13
U8B_ARRAY_3[13]
COMMUNICATION_PROTECTION
168,169
w/r
Communication Protection (enable or disable)
13.14
U8B_ARRAY_3[14]
LOCAL_ADJUST_PROTECTION
168,169
w/r
Local Adjust (enable or disable)
13.15
U8B_ARRAY_3[15]
LOCAL_ADJUST_MODE
170
r
Local Adjust mode (Simple or Complete)
13.16
U8B_ARRAY_3[16]
UNIT_CODE
175,176
w/r
Special Unit Code
13.17
U8B_ARRAY_3[17]
INPUT_UNIT_CODE
179,180
w/r
Special Unit Characterization
13.18
U8B_ARRAY_3[18]
SPECIAL_SENSOR_GAIN
179,180
w/r
Special sensor gain
13.19
U8B_ARRAY_3[19]
SPECIAL_SENSOR_CONNECTION
179,180
w/r
Special sensor connection
13.20
U8B_ARRAY_3[20]
MV_SPECIAL_COLD_JUNCTION
186,187
w/r
Special Cold Junction mode (enable or disable)
14.1
U8B_ARRAY_4[1]
EXE_PV_UPPER_RANGE_VALUE
36
w
Set Primary Variable Upper Range Value
14.2
U8B_ARRAY_4[2]
EXE_PV_LOWER_RANGE_VALUE
37
w
Set Primary Variable Lower Range Value
16.1
FLOAT_ARRAY_1[1]
MEAS_PV_CURR_LEVEL
45
w
Set value of the Trim Primary Variable Current to Dac Zero
16.2
FLOAT_ARRAY_1[2]
UPPER_CAL_POINT_LIMIT
132
r
Upper calibration point limit
16.3
FLOAT_ARRAY_1[3]
LOWER_CAL_POINT_LIMIT
132
r
Lower calibration point limit
16.4
FLOAT_ARRAY_1[4]
CAL_POINT_SPAN
133
w
User span calibration point
16.5
FLOAT_ARRAY_1[5]
PROCESS_VARIABLE
140,184
r
PV Value
16.6
FLOAT_ARRAY_1[6]
SET_POINT
140,146,184
w/r
SP Value
16.7
FLOAT_ARRAY_1[7]
MANIPULATED_VARIABLE
140,147
w/r
MV Value
16.8
FLOAT_ARRAY_1[8]
ERROR
140
r
Error Value
Summary of Contents for HI302
Page 1: ...AGO 02 FOUNDATION H I 3 2 F M E...
Page 4: ...HI302 User s Manual IV...
Page 10: ...General Vision 1 4...
Page 20: ...Installation 2 10...
Page 46: ...Operation 5 12...
Page 57: ...HI302 User s Manual 7 5...
Page 59: ...HI302 User s Manual 7 7...
Page 60: ...An Example of HI302 Usage 7 8...
Page 64: ...Troubleshooting 8 4...