Maintenance of the Split2 Monitor
Inspect the internal face of the curved cover plate and internal surfaces of the sensor head
for dust. If required remove the dust by blowing off with low pressure compressed air.
Inspect the glass lenses of the sensor for dust. If
required clean the dust from the lenses using a
dry foam tipped swab.
Warning:
Do not use a liquid cleaning solution.
Excessive use of cleaning solutions will damage
the matt black paint finish on the inside surfaces
of the sensor head and could penetrate behind
the optical lenses. Failure to comply with this
warning notice will void the product warranty.
Clean external surfaces of the sensor head and
the instrument if required using pre-moistened
lens wipes.
Refit the curved cover plate noting that the foam pad affixed to the underside of the cover
plate is closer to the front edge of the cover plate than the rear edge so that it fits into the
recess in the sensor head body. Secure the cover plate with the three thumb screws, fitting
the top screw first and then the two side screws. Fasten the thumb screws securely to ensure
the sealing of the cover plate to the sensor body.
Refit the components to the front and rear of the sensor head (refer to the exploded diagram
on page 6).
Refit the sensor head to the bracket and secure with the two thumb screws.
Reconnect the sensor head cable to the instrument and perform an auto-zero (refer to page
19) before using the Split2 monitor.
Sensor Lens