Page 16
770-300M Issue E
www.skcltd.com
Setup of the Split2 Monitor for Sampling
2) Check the Battery Status
Always ensure that the battery is fully charged before commencing sampling. The Split2
battery pack has a nominal voltage of 7.2V. A fully charged battery pack will have a voltage of
8.0V or higher. If the battery voltage is below 7.4V it is recommended to recharge the battery
before commencing sampling. Check the battery status as follows:
On the main menu screen press the up and down keys
to highlight the Special Functions menu item, and
press enter to show the new screen.
Run
Review Data
Special Functions
Auto-Zero
On the Special Functions menu screen press enter to
select the System Options menu item and show the
new screen.
System Options
Date/Time
Set Alarm
Cancel
On the System Options menu screen press enter to
select the Extended Options menu item and show the
new screen.
Extended Options
Sample Rate
Erase Memory
Cancel
On the Extended Options menu screen press the up and
down keys to highlight the Battery Status menu item
and press enter to select and show the new screen.
Size Select
Battery Status
Cancel
The Battery Status screen displays the current battery
voltage. Press enter to return to the main menu screen.
Battery Level
Status: 8.05 Volts
3) Set the Particle Size
To set the particle size, on the main menu screen
press the up and down keys to highlight the Special
Functions menu item and press enter to show the new
screen.
Run
Review Data
Special Functions
Auto-Zero
On the Special Functions menu screen use the up and
down keys to highlight the System Options menu item
and press enter to show the new screen.
System Options
Date/Time
Set Alarm
Cancel