Page 30
770-300M Issue E
www.skcltd.com
Battery replacement
To remove the battery pack unscrew the two battery cover screws at the left hand side of the
Split2 monitor case using a size PH1 Phillips screwdriver and remove the battery cover.
Disconnect the battery plug / socket connectors and remove the battery pack from the battery
compartment noting the orientation of the battery pack.
Fit the new battery pack into the battery compartment in the correct orientation and reconnect
the battery plug / socket connectors.
Fit the battery cover in the correct orientation ensuring the foam pad on the rear of the cover
is pressed against the battery pack to secure it in place. Secure the battery cover with the two
screws. Do not apply excessive force when tightening the screws.
Charge the new battery pack fully before use.
Battery disposal
• The EU Battery Directive and equivalent legislation in other countries requires that all
batteries and battery packs are disposed of correctly at the end of their working life. This
means that they must be collected and treated separately from other waste.
• Please ensure that any end-of-life SKC battery packs are collected and recycled or disposed
of correctly.
3) Cleaning the Sensor Optics
It is important to keep the sensor optics of the Split2 monitor clean to ensure the integrity of
the optical sensor.
The condition of the sensor optics should be checked every 48 hours when the Split2 monitor
is used in a 2 to 3 mg/m³ T.W.A. environment, and on a weekly or monthly basis when used in
less contaminated locations.
The required items to clean the sensor optics are as follows -
• Pre-moistened camera lens cleaning wipes.
• Foam tipped swabs.
• Low pressure compressed dry air.
The procedure to clean the sensor optics is
as follows -
Switch off the Split2 monitor and disconnect
the sensor head cable from the monitor.
Remove the sensor head from its bracket
by unfastening the two thumb screws.
Remove all components from the front and
rear of the sensor head.
Unfasten the three thumb screws securing
the curved cover plate on the top of the
sensor head body and carefully remove the
cover from the sensor head body.
Maintenance of the Split2 Monitor