Setup of the Split2 Monitor for Sampling
Note: If performing respirable sampling in Active (pumped) mode perform the auto-zero prior
to fitting the size-selective foam to the IOM inlet on the sensor head.
On the main menu screen press the up and down keys
to highlight the Auto-Zero menu item.
Run
Review Data
Special Functions
Auto-Zero
Press enter and an information screen will briefly
be displayed before the Auto-Zero menu screen is
displayed. Refer to Cleaning the Optics on page 30 for
more information.
AUTO-ZERO
Please clean optics
before auto-zero.
The Auto-Zero menu screen is then displayed. Press
enter to select the Auto-Zero menu item and initiate
the Auto-Zero sequence. The Auto-Zero status screen
is then displayed.
Auto-Zero
Cancel
The sequence takes 25 seconds and an on-screen
countdown timer indicates progress. At the end of the
sequence the display will automatically return to the
main menu screen.
Auto-Zeroing
Please Wait
022
On completion of the auto-zero, if performing Active sampling switch off the air sample pump
and disconnect the zeroing accessory from the IOM inlet on the sensor head.