www.skcltd.com
770-300M Issue E
Page 27
Downloading the Sample Data to a PC
Refer also to the DustComm Pro software manual for detailed instructions on the installation
and use of the software.
The following is a summary of the process of downloading the sample data.
Connect the Split2 monitor to the serial port on the PC using the supplied RS-232 data cable.
Ensure that both connections are secure as an intermittent connection will disrupt the data
transmission.
A USB to serial adapter cable (not SKC supply) is required if the PC does not have a serial
interface port. The USB to serial adapter must be configured in Windows Device Manager
such that its COM port number is set to an available number in the range COM1 to COM4.
Run the DustComm Pro software. Select the Unit menu option on the DustComm Pro window
and select Properties.
On the Properties window ensure that the Unit Type is set to “Split 2” from the drop down list.
Ensure that the Download Port is set to the correct COM port number on the PC from the drop
down list. Click the OK button to close the Properties window.
Select the Unit menu option on the DustComm Pro window and select Download. A progress
window will open with the message “Awaiting connection...”.
On the Split2 monitor, from the main menu screen
press the up and down keys to highlight the Review
Data menu item. Press enter to display the Review
Data menu screen.
Run
Review Data
Special Functions
Auto-Zero
On the Review Data screen press the up and down keys
to highlight the Download menu item. Press enter to
display the Download menu screen.
Statistics
Download
Cancel
On the Download menu screen press enter to select
the To Data Collector menu item and initiate the data
download.
To Data Collector
Cancel
The screen will show a Transmitting message and
then revert to the Download menu screen indicating
that the data transfer is complete.
Transmitting
Please Wait
On the DustComm Pro software progress window the message “Connection established,
downloading data...” will be displayed and a progress bar will show the data transfer
proceeding.
On completion of the data transfer the progress window will show the message “The download
was successful!”. Press the OK button on the window to close it.