Instruction Manual StackGuard
34
10266E/1
Action Display
(example)
Remarks
18.
Check value
updated
If the display shows “out
of tolerance” or
“Measurement” error, check
whether the checking rod has
really been removed.
Otherwise there is a fault. If
that is the case, contact your
service partner.
19.
p
* RECALIBR. *
* *
20.
8 x
n
* ADJUSTMENT *
* INFO *
21.
p
Recal1 -0.5%
24.11.04 16:05h
Enter the value of the devia-
tion in the maintenance pro-
tocol
22.
q
+
p
(simultane-
ously)
0.007 PLA 4
Unit in normal mode.
23.
Rebuild the unit to the initial state.
24.
Enter the work in the maintenance protocol.
If the difference between nominal and actual values is too great when
recalibrating, “
Recalibr. out of tolerance
” is displayed. If this happens,
check the following:
Action Bemerkung
1.
Check whether the measured value
corresponds to the value on the control unit.
→
Point 3
2.
Check the control unit for cleanliness.
→
Point 6
3.
Has the minimum warm-up time of two hours
been observed?
4.
Repeat the recalibration.
Let the checking rod
cool down before trying
again!
5.
If the recalibration is still not OK, contact
your service partner.
What to do if value
is “out of
tolerance”?
Check list for
recalibration