Instruction Manual StackGuard
26
10266E/1
4.4
Setting the scaling and the unit
The unit can be switched to a different unit by the PLA.
For the display in mg/m
3
the corresponding scale factor must be defined using
a reference measurement procedure!
The unit „PLA“ can be set only if the scaling factor has been set to 1.000 (
→
Reference Handbook)
Action Display
(example)
Remarks
1.
Activate service
mode.
* SENSORCHECK *
→
2.
3 x
n
* CONFIGURE *
3.
3 x
q
> Scaling <
1.000
4.
Activate editing mode
with
2
.
Scaling
> 1.000 <
5.
Set value with
o
/
n
.
Scaling
> 1.000 <
Values to be set:
PLA = 1.000
6.
Confirm value with
2
.
> Scaling <
1.000
7.
p
> Unit <
PLA
8.
Activate editing mode
with
2
.
Unit
> PLA <
9.
Letter selection
with
o
/
n
.
Unit
> mLA <
Small letters and spe-
cial characters are also
possible by using the
o
/
n
keys.
10.
Change position
with
q
/
p
.
Unit
> mgA <
11.
Confirm selection
with
2
.
> Unit <
mg/m3
12.
q
+
p
(together)
0.007 mg/m3 5
4.5
Setting the relay functions
The control unit possesses two relay outputs (
→
functions are freely configurable. Several functions can be assigned
simultaneously to a given relay. This relay then becomes active whenever one
of the configured functions is active (OR operation).