Instruction Manual StackGuard
10266E/1
33
Action Display
(example)
Remarks
8.
Wait until the measuring cell has been
vented by the purge air and then check
the reading.
The reading must now reduce to less than
0.001 PLA.
If this is not the case, the optics of the
measuring cell to be cleaned. Contact your
service partner.
9.
Loosen the knurled nut (A) on the check-
ing rod and then press in the pin (B).
Now you can pull the checking rod out of
the protective sleeve
Insert the checking rod into the photome-
ter. The pin (C) must snap into the groove
(figure).
10.
Turn the checking rod clockwise to the
stop and re-tighten the knurled nut (A).
11.
Wait until a stabile
reading is dis-
played
- Reading -
7.52
If the value differs from the
nominal value only a little,
you can continue at point 26.
12.
p
> Recalibr. <
Retain
Recalibration is carried out
13.
2
Recalibr.
> Retain <
14.
p
Recalibr.
> Adapt <
15.
Initiate the re-
calibration by
pressing the
2
key.
Recalibr.
running ...
16.
Wait until the
StackGuard has
determined the
calibration factors
and the display
changes.
Remove
checking rod!
Remove the checking rod
If “out of tolerance” ap-
pears in the display, consult
the following section.
17.
p
Acquire
check value...
The values of the internal
control unit are updated