Laboratory centrifuge Sigma 1-7
6 Using the centrifuge
30 / 68
Version 08/2018, Rev. 1.7 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0703402
Pos : 105 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /061-0020-0020 Ei nsetz en von R ot oren und Zubehör @ 27\m od_1405320317739_68. docx @ 193725 @ 3 @ 1
6.2.2
Installation of rotors and accessories
Pos : 106 /100 Sigma/110 BA Z entrif ugen Sigma (Proj ekt e)/ 1-16_1- 16K_1-7/061 Betri eb/061-0020-0020- 0010 Ei ns etz en ei nes R ot ors 1- 16_1-16K_1-7 @ 28\mod_1405321059727_68. docx @ 196007 @ 4 @ 1
6.2.2.1
Installation of a rotor
• Open the centrifuge lid by pressing the lid key.
• Loosen the rotor tie-down screw by turning it anti-clockwise, but do not
remove it.
• Lower the rotor with its central bore straight down onto the motor shaft.
• Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 3 Nm. In doing so, hold the rotor at its outer rim.
• Follow the safety instructions and hazard warnings (see chapter 3 -
WARNING
Once a day or after 20 cycles, the rotor tie-down screw must be loosened
by some turns, and the rotor must be lifted and fastened again. This
ensures a proper connection between the rotor and the motor shaft.
NOTE
Rotors can be used without a cover. This leads, however, to higher levels
of noise and temperatures when running.
Removal of a rotor
• Loosen the rotor tie-down screw by turning it anti-clockwise and remove
the rotor.