Laboratory centrifuge Sigma 1-7
5 Set-up and connection
28 / 68
Version 08/2018, Rev. 1.7 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0703402
5.2.2
Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp B fuses that are to
be provided by the customer.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.