Laboratory centrifuge Sigma 1-7
3 Safety
24 / 68
Version 08/2018, Rev. 1.7 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0703402
3.7.2
Standstill monitoring system
Opening of the centrifuge lid is only possible if the rotor is at a standstill.
This standstill is checked by the microprocessor.
Pos : 63 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 64 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0032 Syst emk ontroll e Spincontr ol Basic+ Eas y+Univ ersal @ 27\mod_1405319041777_68. docx @ 192227 @ 3 @ 1
3.7.3
System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. The system continuously performs a self-check and
identifies malfunctions. Malfunctions are indicated by error messages with a
number in the speed/rcf display (see chapter 7.2 - "Table of error codes").
Pos : 65 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 66 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0041 Sc hutzleit erpr üf ung ohne Potenti alausgleic hssc hraube @ 27\m od_1405319044593_68. docx @ 192255 @ 3 @ 1
3.7.4
Earth conductor check
An earth conductor check can be carried out by authorised and specialised
personnel using a suitable measuring instrument. Please contact the Sigma
service department (see chapter 7.3 - "Service contact").
Pos : 67 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 68 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 69 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0080 Ver halt en bei Gefahren und Unfäll en @ 27\mod_1405319051188_68. docx @ 192325 @ 2 @ 1
→
3.8
Measures in the event of hazards and accidents
DANGER
• If an emergency arises, switch off the centrifuge immediately!
• If in doubt, call the emergency doctor!
Pos : 70 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 71 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 72 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0100 R estrisiken @ 27\mod_1405319052885_68. docx @ 192339 @ 2 @ 1
→
3.9
Remaining hazards
The centrifuge was built in accordance with the state of the art and in
compliance with the generally recognized safety rules. However, danger to
life and limb of the operator, or of third parties, or impairments of the unit or
other material assets cannot be completely excluded when the unit is being
used.
• Use the unit only for the purpose that it was originally intended for (see
chapter 1.2 - "Intended use").
• Use the unit only if it is in a perfect running state.
• Immediately eliminate any problems that can affect safety.