8
TECHNICAL DATA
68
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Operating conditions
Temperature
Temperature
Temperature
Temperature
Product
-40
…
+240
°
C / -40
…
+465
°
F
Ambient
Non-Ex: -40
…
+85
°
C / -40
…
+185
°
F
Ex: -40
…
+60
°
C / -40
…
+140
°
F
Storage
-50
…
+85
°
C / -58
…
+185
°
F
Pressure
Pressure
Pressure
Pressure
Product
Max. 100 bar / 1450 psi; Information on higher pressures on request.
Ambient
Atmosphere
Chemical properties
Chemical properties
Chemical properties
Chemical properties
Density
Taken into consideration when sizing.
Viscosity
< 10 cP
Reynold's number
10000...2300000
Recommended flow velocities
Recommended flow velocities
Recommended flow velocities
Recommended flow velocities
Liquids
0.3
…
7 m/s / 0.98
…
23 ft/s (optional up to 10 m/s / 32.8 ft/s taking
cavitation into account)
Gases and steams
2.0
…
80 m/s / 6.6
…
262.5 ft/s
DN15: 3.0
…
45 m/s / 9.8
…
148 ft/s; DN25: 2.0
…
70 m/s / 6.6
…
230 ft/s
For detailed information, see chapter "Flow tables".
Other conditions
Other conditions
Other conditions
Other conditions
Protection category
IP 66/67
Installation conditions
Inlet run
≥
20 x DN (without disturbing flow, after pipe narrowing, after a single
90
°
bend)
≥
30 x DN (after a double bend 2x90
°
)
≥
40 x DN (after a double three-dimensional bend 2x90
°
)
≥
50 x DN (after control valves)
≥
2 DN before flow straightener;
≥
8 DN after flow straightener
(specified values apply only to original
≥
20 DN inlet run)
Outlet run
≥
5 x DN
Dimensions and weights
For detailed information refer to chapter "Dimensions and weights".
Materials
Measuring sensor and process
connections
Standard: 1.4404/316L
Option: Hastelloy
®
C-22 on request
Converter housing
Die-cast Aluminium
Pressure sensor gasket
Standard: FPM
Option: FFKM
Measuring tube gasket
Standard: 1.4435/316L
Option: Hastelloy
®
C-276
Selection depends on measuring sensor material/medium.
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 68 Monday, June 22, 2009 8:17 AM