SERVICE
7
63
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02 A5E02100423
Service
7.1 Replacing signal converter / LCD display
The signal converter has to be replaced by a converter of the same type. The following
parameters have to be observed for it:
•
The item number must match: 2.143670.xxx
•
The software version
2
must match.
•
The basic version has no software identification
•
The gas version is marked: "gas"
•
The steam version is marked: "steam"
The following steps are to be performed:
•
Switch off the power supply.
•
Unscrew the front cover.
•
Detach the display from the spacer pins.
•
Detach the display cable
1
.
•
Detach the measuring sensor cable
3
.
•
Loosen the three fixing screws
5
(Phillips).
•
Pull out signal converter.
•
Insert new signal converter.
•
Tighten the three screws.
•
Attach measuring sensor cable
3
(Caution: ensure that the measuring sensor cables are not
on top of the spacer pins
4
May be damaged).
•
Attach display cable
1
.
•
Attach display in desired position, apply even pressure to the entire surface.
•
Screw on cover by hand.
Figure 7-1: Board, example
1
Connector for display
2
Version number, software identification
3
Connector for measuring sensor
4
Spacer pins
5
Fixing screw (Phillips)
MA_Siemens_Sitrans_FX300_R02_en_PRT.book Page 63 Monday, June 22, 2009 8:17 AM