3
INSTALLATION
30
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02 A5E02100423
3.6.3 Installing devices in sandwich design
•
Push the first bolt
3
through the hole
5
of both flanges.
•
Screw on the nuts and washers to both ends of the bolt
3
but do not tighten them.
•
Install the second bolt through the holes
4
.
•
Place the measuring sensor
1
between the two flanges.
•
Insert the gaskets
6
between measuring sensor
1
and flanges and align them.
•
Check that the flange is concentric.
•
Install the remaining bolts, washers and nuts. Do not yet tighten the nuts.
•
Turn the centring ring
2
in a counter-clockwise direction and align the device.
•
Check that the gaskets
6
are concentric; they must not protrude into the pipe cross-section.
•
Now tighten all nuts bit by bit alternately across the diagonal.
Figure 3-13: Installation using centering ring
1
Measuring sensor
2
Centering ring
3
Bolts with fixing nuts
4
Hole
5
Hole
6
Gasket