ELECTRICAL CONNECTIONS
4
37
SITRANS FX300
www.siemens.com/flow
06/2009 - SFIDK.PS.050.F2.02 A5E02100423
4.5 Protection category
The measuring device meets all requirements of protection category IP66/67.
Therefore it is essential to observe the following points:
•
Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
•
The electrical cables used must be undamaged and must comply with regulations.
•
The cables must be laid with a loop
1
upstream of the measuring device to prevent water
from getting into the housing.
•
The cable feedthroughs
2
must be tightened.
•
Align the measuring device so that the cable feedthrough is never facing up
3
.
•
Close the unused cable feedthroughs using blind plugs
4
.
•
Do not remove the required cable bushing from the cable feedthrough.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection
category must be ensured again.
Figure 4-6: Cable feedthrough