Functions
7.1 Setting Liquid Parameters
FUH1010 IP65 NEMA 4X Precision Volume
Operating Instructions, 01/2013, A5E32122003-AA
85
Setting the Pressure Slope
The flow meter measures and reports the flowing liquid’s sonic velocity (Vs) and its related
pressure (P). Since a pressure change will affect the sonic velocity of the liquid, a method to
balance the measured sonic velocity to a fixed reference temperature (15.6ºC/60ºF) default)
is provided. The Pressure Slope Factor default is 0.030 (m/s per PSI). This pressure slope
factor helps maintain an accurate liquid identification as the LiquIdent factor and liquid
temperature vary.
The flow meter has the ability to accept a 4-20 mA input for pressure. If used, the flow meter
will process pressure data based on the analog input. This is recommended for pipelines that
have wide pressure variances.
To calculate a liquid's Pressure Slope Factor:
1.
Establish the maximum (Pmax) and minimum (Pmin) pressures for all liquids within the
Liquid Class. For each liquid, note the measured sonic velocity at Pmax and the sonic
velocity at Pmin (in-meters-per-second).
2.
Use the following formula to calculate the Pressure Slope Factors with the Liquid Class:
3UHVVXUH6ORSH)DFWRU 9
V
#
3
PLQ
9
V
#
3
PD[
3
PLQ
3
PD[
3.
Calculate the average of all the Pressure Slope Factors within the Liquid Class.
To enter the Pressure Slope Factor
1.
From the [Liquid Table] menu cell, press <Up/Down> Arrow> to scroll down to [Pressure
Slope].
2.
Press <Right Arrow> to enable numeric entry.
3.
Use the keypad numeric keys to enter the desired Pressure Slope Factor.
4.
To enter the Pressure Slope Factor press <ENTER>.
5.
This moves the highlight down to [Base Temperature].
Setting the Base Reference Temperature
The flow meter uses a reference Base Temperature of 15.6ºC (60ºF) for the LiquIdent
calculation and Standard Volume compensation. Any temperature can be entered and used
as the reference Base Temperature (e.g., 20ºC/68ºF).
To enter the Base temperature:
1.
From the [Liquid Table] menu cell, press <Up/Down Arrow> to scroll down to [Base
Temperature].
2.
Press <Right> to enable numeric entry.
3.
Use the keypad numeric keys to enter the desired Base Temperature.
4.
To enter the Base Temperature press <ENTER>.
Summary of Contents for SITRANS FUH1010
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ...21614 C 1011NFPS 7 ...
Page 232: ...1011NFPS 7 C 21614 ...
Page 233: ...1011NFPS 7 C 21614 ...
Page 234: ...21614 C 1011NFPS 7 ...
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...21614 C ...
Page 241: ......
Page 242: ......
Page 243: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 244: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 245: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 246: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 263: ...Index FUH1010 IP65 NEMA 4X Precision Volume 202 Operating Instructions 01 2013 A5E32122003 AA ...