Troubleshooting
10.1 Troubleshooting
FUH1010 IP65 NEMA 4X Precision Volume
140
Operating Instructions, 01/2013, A5E32122003-AA
Error or Message
Probable Cause
Solution
Invalid Setup
During the Initial Makeup the system
detects invalid transducer spacing,
erroneous liquid pipe parameters, or some
other factor that prevents it from completing
the Initial Makeup.
This may be due to one of the following:
An out-of-range data entry.
An invalid condition (e.g., overlapping
transducers in Reflect Mode).
In Reflect Mode the flow meter detects that the
pipe wall signal may infringe upon the signal.
Press <ENTER>, <Up Arrow>, <Down Arrow>,
or <Left Arrow> to abort install routine.
Continue programming other site data in
anticipation of resolving the difficulty later. Call
technical support for help, if necessary.
Low Signal - Press
<ENTER>
During the Initial Makeup the flow meter
decides that the level of the receive signal is
insufficient for proper operation.
Some reasons for low signal are:
Invoking [Install Complete] on an empty pipe.
Coupling compound insufficient; not applied or
evaporated.
A disconnected or broken transducer cable.
The pipe needs to be conditioned at the
mounting location.
Flush out large air bubbles.
The sensor cables are defective or not
connected to the correct channel.
The Set Empty routine performed when pipe
was NOT actually empty.
If you locate and correct the improper condition
immediately, press <ENTER> to resume the
installation procedure. Otherwise, press the <Left
Arrow> to abort the installation and conduct a
thorough investigation.
Detection Fault
If it appears that the flow meter cannot
complete an Initial Makeup it means that the
pipe and/or liquid conditions do not permit a
receive signal that meets the flow detection
standards. The system will not operate.
Attempt to improve operating conditions by
reinstalling the sensors at a different spacing
offset, or even at a different location on the pipe.
Switching from Reflect to Direct Mount may solve
the problem. However, operation may not be
possible if there is poor liquid, pipe wall sonic
conductivity or extreme liquid aeration. Call
technical support for help, if necessary.
Note
If you receive a Detection Fault message, it is strongly recommended that the Technical
Service Department (
http://www.automation.siemens.com/partner
) be contacted.
Summary of Contents for SITRANS FUH1010
Page 194: ......
Page 195: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ...21614 C 1011NFPS 7 ...
Page 232: ...1011NFPS 7 C 21614 ...
Page 233: ...1011NFPS 7 C 21614 ...
Page 234: ...21614 C 1011NFPS 7 ...
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ...21614 C ...
Page 241: ......
Page 242: ......
Page 243: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 244: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 245: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 246: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 263: ...Index FUH1010 IP65 NEMA 4X Precision Volume 202 Operating Instructions 01 2013 A5E32122003 AA ...