Parameters
SITRANS F M TRANSMAG 2
88
Operating Instructions, 11/2009, A5E00102775-07
6.5.3
Volt. Uel1
Control value for voltage between
electrode 1 and medium
6.5.4
Volt. Uel2
Control value for voltage between
electrode 2 and medium
6.5.5
Flow % Range
1,2
Actual flow in % of set full scale value
6.5.6
Sampl. Frequ.
Actual measuring frequency
(approx. 2 x
magnetic
frequency)
6.5.7
Selftest
0 %
FS
Display of internal control values of
measuring electronics. Run self-test to
update (menu 2.4.1)
Zero
Gain
(result display of the
last self-test)
(xxxxxxxx)
(current status
active/inactive)
Triggered when
selecting menu 2.4.1
6.5.8
Service Info
Coded service information
(Info)
6.5.9
Digital Input
2
Status of the digital input under
consideration of the set signal type
(actual status
active/inactive)
6.6
Zero Trim
6.6.1
Zero Flow Cor.
Correction value for the zero point of the
flow measurement in m/s
-1 ... +1 m/s
6.6.2
Start Correct.
Start of zero calibration; The medium
must be at a standstill during
calibration!
Cancel
•
Cancel
•
Start
6.7
Trim Parameter Values in 6.7.2.4 can only be written if no
SmartPlug data can be read
6.7.1
Cal.factor
User-specific calibration factor; the
measured flow value is multiplied by this
factor before output (affects all outputs,
totalizers and display)
1.0
0.5 ... 1.5
6.7.2
CFH
5
CFR
5
Calibration value for hydraulic reference
point at 1m/s
Calibration value for reference path
400
100
6.7.3
ZPH
5
Calibration value for
hydraulic zero point
0