SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
23
Connecting
5
5.1
General Information
This chapter describes how to wire up the device.
Step 1 (Connecting power supply (Page 25)) and step 2 (Connecting signal cables
(Page 26)) must be carried out for compact as well as remote versions.
Step 3 (Connecting terminal box and transmitter (Page 27)) is only relevant for remote
versions.
The chapter only describes wiring of devices with HART. In order to wire up devices with
Profibus PA, refer to the Appendix:
1
2
3
4
5
6
7
①
Magnetic current line
②
Terminal box for power supply and signal lines
③
Potential equalization
④
Signal line
⑤
Power supply
⑥
Electrode line
⑦
Terminal box for magnetic current and electrode line
Figure 5-1
Overview, Electrical connections