7.5 Function Group Device Outputs
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
55
scaled to the 0 to "Split value" range. If the flow value exceeds this subdomain, the output
current is scaled to the 0 to URV range.
Note
The relay contact is open in the "no function" setting.
● Signal type, menu item 4.3.2
You can set the function of the relay as follows in this menu item:
1.
When the relay need to be closed in the event of an alarm:
Flow in forward direction
Limit value message
2.
When the relay need to be open in the event of an alarm:
Flow in forward direction
Limit value message
Table 7- 3
Relay status
Setting
No alarm
Reverse flow
No limit signalling
Alarm message
Forward flow
Limit signalling
1. Contact closes
2. Contact opens
①
Electronic fuse
②
External supply
Figure 7-8
Digital output 2