Profibus Communication
B.8 Device Database File (GSD)
SITRANS F M TRANSMAG 2
Operating Instructions, 11/2009, A5E00102775-07
101
Example
The measured values "Totalizer net" and "Totalizer forwards" should be transmitted. Then
the following identifier combination must be specified:
● Free Place (for flow)
● long Identifier (for totalizer net)
● long Identifier (for totalizer forwards)
No identifier for the totalizer reverse is necessary here because the value of the totalizer
forwards is the last measured value to be transmitted.
The following files (GSD and bitmap) can be downloaded from the internet using the address
http://www.ad.siemens.de/csi_e/gsd
These files are also implemented in the device install files for SIMATIC PDM software.
GSD files (version 31.07.2001)
#Profibus_DP
GSD_Revision = 3
Vendor_Name = "SIEMENS AG"
Model_Name = "SITRANS FM"
Revision = "Revision 01"
Ident_Number = 0x80C4
Protocol_Ident = 0
Station_Type = 0
FMS_supp = 0
Hardware_Release = "A01"
Software_Release = "Z01"
Bitmap_Device = "SIE80C4n"
31.25_supp = 1
45.45_supp = 1
93.75_supp = 1
MaxTsdr_31.25 = 100
MaxTsdr_45.45 = 250
MaxTsdr_93.75 = 1000
Redundancy = 0
Repeater_Ctrl_Sig = 0
24V_Pins = 0
Freeze_Mode_supp = 0
Sync_Mode_supp = 0
Auto_Baud_supp = 0