Connecting
5.4 Step 3: Connecting terminal box and transmitter
SITRANS F M TRANSMAG 2
28
Operating Instructions, 11/2009, A5E00102775-07
Connecting
The sensor cable shield is connected to the transmitter and sensor as follows:
1.
Bend the cable shield back over the clamping piece of the cable gland.
2.
Push the clamping piece with the sensor cable into the threaded bush of the cable gland
turning it slightly to the right.
3.
Tighten the lock nut on the threaded bushing until the cable is connected tightly (IP67)
①
Electrode line
②
Magnetic current line
Figure 5-4
Signal transmission from the sensor to the transmitter