
MAG 8000 & MAG 8000 CT
44
SFIDK.PS.026.D9.02
Transmitter
Installation MAG 8000
Integral (compact) or remote with factory mounted cable in
5, 10, 20 or 30 m lengths with IP68/NEMA 6P connectors.
Connection is made at the transmitter bottom.
Installation MAG 8000 CT
Integral (compact) or remote with factory mounted cable in 5 or
10 m lengths with IP68/NEMA 6P connectors.
Connection is made at the transmitter bottom.
MI-001 verified meters only Integral (compact) mounted.
Enclosure
Stainless steel top housing (AISI 316) and coated brass bottom.
Remote wall mount bracket in stainless steel (AISI 304).
Cable entries
2 x M20 (one gland for one cable of size 6 … 8 mm
(0.02 ... 0.026 ft) is included in the standard delivery)
Display and key
Display with 8 digits for main information.
Index, menu and status symbols for dedicated information
Key for toggling through the information and reset customer
totalizer and call-up function
Selectable default information and accessible menus:
- Operator
- Meter
- Service
- Data Logger
- Statistic and leakage (only Advanced version)
- Revenue and Tariffs (only Advanced version)
Totalized information can be displayed with 1, 2, 3 decimals or
automatic adjustment for maximum resolution
Flow unit MAG 8000
Europe std.
Volume in m
3
and flow rate in m
3
/h
US std.
Volume in Gallon and flow rate in GPM
Australian std.
Volume in Ml and flow rate as Ml/d
Other units selectable:
Volume: m
3
x 100, l x 100, G x 100, G x 1000, MG,
CF x 100, CF x 1000, AF, Al, kl
Flow: m
3
/min, m
3
/d, l/s, l/min, GPS, GPH, GPD, MGD, CFS,
CFM, CFH
Other units are ordered from factory or manually configured onsite
by sticking a label on the display and changing the scaling factors
Flow unit MAG 8000 CT
Europe std.
Volume in m
3
and flow rate in m
3
/h
Digital output MAG 8000
2 passive outputs (MOS), individual galvanically isolated
Maximum load
±
35 V DC, 50 mA short circuit protected
Output A function
Programmable as pulse volume
– forward – reverse – forward/net
– reverse/net
Output B function
Programmable as pulse volume (like output A), alarm or call-up
Output
Max. pulse rate of 50 Hz (only Basic version) and 100 Hz (only
Advanced version), pulse width of 5, 10, 50, 100, 500 ms
Digital output MAG 8000 CT
2 passive outputs (MOS), individual galvanically isolated
Maximum load ± 35 V DC, 50 mA short circuit protected
Output A function
Programmable as pulse volume
– forward
Output B function
– Alarm
Output
Max. pulse rate of 50 Hz (only Basic version) and 100 Hz (only
Advanced version), pulse width of 5, 10, 50, 100, 500 ms
8. Technical data
(continued)
Summary of Contents for SITRANS F M MAG 8000
Page 60: ...MAG 8000 MAG 8000 CT 60 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 61: ...MAG 8000 MAG 8000 CT 61 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 62: ...MAG 8000 MAG 8000 CT 62 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 63: ...MAG 8000 MAG 8000 CT 63 SFIDK PS 026 D9 02 10 Approvals and certificates ...