MAG 8000 & MAG 8000 CT
21
SFIDK.PS.026.D9.02
4. Commissioning
Net flow output
The MAG 8000 has a special net pulse output that includes bi-directional flow calculations.
The example shows that over time, the net pulse output indicates the bi-directional totalizer as
calculated internally.
The same principle applies for forward and reverse flow calculation. By changing the status of
the pulse output, the internal pulse calculator will be reset.
Net totalizer in
meter display
(Bi-directional)
Pulse output forward
Uni-directional mode
Volume [m
3
]
Pulse output net forward
Bi-directional mode
Volume [m
3
]
Flow
Volume [m
3
]
Delivered
volume
Internal
calculation
Total accounted volume
[m
3
]
F
orward/
R
everse
Delivered
volume
Internal
calculation
0
-
0
0
0
10
-
10
0
10
−
2
-
0
−
12
0
18
-
20
−
12+20=
8
18F
30F
18F
FT ID
Meter
Display Parameter/data type
Factory settings
Data range
number
version view
Fix parameter or meter data that
not
are
changeable
1
All
M1
Application identifier
Indetity
Max. 14 characters. Only numbers
are visible on the display
2
All
-
Application location
Location
Max. 14 characters
3
All
M3
Module type
MLFB depended
Basic or Advanced
4
All
M4
Software version
x.xxPxx (x.xxPx.x)
5
All
-
Sensor size
Sensor related
DN 25..600(1"..24")
6
All
-
Vendor name
Siemens
Siemens
8
All
-
Totalizer unit
MLFB depended
Max. 10 characters
9
All
-
Flowrate unit
MLFB depended
Max. 10 characters
10
All
-
Qn
(Q3)
Sensor related
0 to 1*10^9
11
All
-
Product code number
7ME6810XXXXXXXXX
7ME681 or 7ME682 XXXXXXX
12
All
-
Serial number
XXXXXXNXXX
100
All
M2
Actual date and time
PS3 production date and time
year-month-day T
hours:minutes:seconds
101
All
O1
Totalizer 1
0
0 to +-2*10^9
102
All
O2
Totalizer 2
0
0 to +-2*10^9
103
All
O5
Customer totalizer 3
0
0 to +-2*10^9
104
All
O5
Reset customer totalizer 3
No
Yes/no
4.3 Parameter list
MAG 8000 is delivered with factory settings that are not stored as default values. Because defaults
values are not present in the meter, an automatic return to factory values is not possible.
The default settings are available at
www.siemens.com/flow
. Navigate to Tools & Downloads
under MAG 8000.
Visible display information is indicated in the table by menu and index number. Remember to
enable displayed menus FT130.
The abbreviations used in the display menu table are: Operator menu = O, Meter menu = M,
Service menu = Se, Data Logger menu = L, Statistic menu = St, Revenue menu = R.
Summary of Contents for SITRANS F M MAG 8000
Page 60: ...MAG 8000 MAG 8000 CT 60 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 61: ...MAG 8000 MAG 8000 CT 61 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 62: ...MAG 8000 MAG 8000 CT 62 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 63: ...MAG 8000 MAG 8000 CT 63 SFIDK PS 026 D9 02 10 Approvals and certificates ...