MAG 8000 & MAG 8000 CT
41
SFIDK.PS.026.D9.02
7. Service
7.1.1 Faults codes
(continued)
Faults
Alarm description
Alarm reasons
codes
Warning
Indication of wrong setting or application issues
5
Low power
Battery voltage is below 2 volts or battery capacity is below the low
power % limit (set as customer parameter FT206). When low
voltage is detected, flow measurement and communication will stop,
but the display will remain active as long as some power is
available. The alarm remains active until the fault condition is
resolved.
-> Check the calculated battery capacity (FT510) versus the
battery alarm limit (FT206) and replace batteries as necessary.
6
Flow overflow
Flow rate through the meter is 25% above Qn (Q3) and the meter
stops counting at 125% flow. The alarm remains active until the
fault condition is resolved.
-> Check the sizing of the meter for the current installation.
7 & 8
Pulse 1 & 2 overflow
The pulse rate is higher than the pulse output can deliver.
Reminder: the
basic version
is limited to 50 Hz maximum; the
advanced version
is limited to 100 Hz maximum. The alarm
remains active until the output pulse rate drops below the maxi-
mum pulse rate.
-> Change volume per pulse to a higher value - see section 8
“Technical data“ for pulse selection.
9
Consumption warning
The data logged consumption on totalizer 1 has exceed the too low
or too high consumption limit. The alarm remains active until it is
manually reset by FT209.
-> Check data logger values and consumption limit.
L
Leakage
The lowest flow rate or volume during leakage period has
(Advanced version
exceeded the leakage detection settings. The alarm is active until
only)
it is manually reset by FT208. An alarm reset (similar to leakage
period reset FT820) will reset information for the selected function.
-> Check setting and pipe installation.
E
Empty pipe
The measured electrode impedance is above the empty pipe level
(FT540 & FT541 & FT334). The alarm remains active until the fault
condition is resolved.
-> Make sure the sensor is filled with water.
C
High conductivity
The electrode impedance is below the low media impedance
(FT542) - meaning the water has a high conductivity. The alarm is
active until the resistance in the water is above the low media
alarm limit.
d
High flow rate
The flow rate is greater than the flow alarm limit (FT553).
The alarm remains active until the flow rate drops below the flow
alarm limit.
Warning faults cover operating conditions where settings or monitored values exceed configured
limits.
Some of the warnings require a manual reset before they are removed from the alarm list.
Note
A reset of the fault log (FT204) also resets all alarms. Once reset, only active alarms become
visible again.
Summary of Contents for SITRANS F M MAG 8000
Page 60: ...MAG 8000 MAG 8000 CT 60 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 61: ...MAG 8000 MAG 8000 CT 61 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 62: ...MAG 8000 MAG 8000 CT 62 SFIDK PS 026 D9 02 10 Approvals and certificates ...
Page 63: ...MAG 8000 MAG 8000 CT 63 SFIDK PS 026 D9 02 10 Approvals and certificates ...